sudo find / -name pysmu
/home/mackst/.local/bin/pysmu
/home/mackst/.local/lib/python3.6/site-packages/pysmu-1.0.2-py3.6-linux-x86_64.egg/EGG-INFO/scripts/pysmu
/home/mackst/.local/lib/python3.6/site-packages/pysmu-1.0.2-py3.6-linux-x86_64.egg/pysmu
Bei Python 3 findet der Interpreter pysmu unter ...
python3
>>> import pysmu
>>> print(pysmu.__file__)
/usr/local/lib/python3.6/dist-packages/pysmu-1.0.2-py3.6-linux-x86_64.egg/pysmu/__init__.py
6.6.20
Pythonbeispiele aus wiki.analog.com/university/tools/python-tutorial/table-of-contents
Direct control over ADALM1000 functionality can be accomplished using the implemented control
transfers, a synchronous and slow communication endpoint accessible regardless of the
configuration of the device.
Note that the higher level functions libpysmu implements on top of LIBSMU make use of many of
the control transfers to configure device state during normal operation. As such, using these
low level control transfers while streaming data may not produce the expected results.
The ALM1000 implements many control transfers, including the following:
0x17 - read a number of bytes from the ADM1177 hot-swap controller
0x50 - Set a GPIO pin low
0x51 - Set a GPIO pin high
0x91 - Get a GPIO input pin value
0x53 - Set device Mode
0x59 - Set potentiometer state
M1000 microcontroller pin mappings are described below.
IO Name Net Name Net Description Pin ID
——- ——-- ————— ——
PA0 PIO0 user DIO 0-220Ω 0
PA1 PIO1 user DIO 1-220Ω 1
PA2 PIO2 user DIO 2-220Ω 2
PA3 PIO3 user DIO 3-220Ω 3
PA4 PIO0 user DIO 0-470Ω 4
PA5 PIO1 user DIO 1-470Ω 5
PA6 PIO2 user DIO 2-470Ω 6
PA7 PIO3 user DIO 3-470Ω 7
PB0 UA11-IN0 50Ω from ChA to 2v5 32
PB1 UA11-IN1 50Ω from ChA to GND 33
PB2 UA11-IN2 ChA close voltage sense loop 34
PB3 UA11-IN3 ChA connect output 35
PB5 UB11-IN0 50Ω from ChB to 2v5 37
PB6 UB11-IN1 50Ω from ChB to GND 38
PB7 UB11-IN2 ChB close voltage sense loop 39
PB8 UB11-IN3 ChB connect output 40
PB17 PWR_ENABLE turn on power for analog components 49
PB19 SWMODE-A ChA switch SVMI - SIMV 51
PB20 SWMODE-B ChB switch SVMI - SIMV 52
Examples:
#
import libpysmu
import pysmu.py
devx = Smu()
DevID = devx.serials[0] # device ID for 1st M1000 in list
# set PIO1 high
devx.ctrl_transfer(DevID, 0x40, 0x51, 1, 0, 0, 0, 100)
# set PIO1 low
devx.ctrl_transfer(DevID, 0x40, 0x50, 1, 0, 0, 0, 100)
# get state of PIO0
print devx.ctrl_transfer(DevID, 0xc0, 0x91, 0, 0, 0, 1, 100)
# get state of PIO1
print devx.ctrl_transfer(DevID, 0xc0, 0x91, 1, 0, 0, 1, 100)
# set CHA 2.5 V switch to open
devx.ctrl_transfer(DevID, 0x40, 0x51, 32, 0, 0, 0, 100)
# set CHA GND switch to open
devx.ctrl_transfer(DevID, 0x40, 0x51, 33, 0, 0, 0, 100)
# set CHB 2.5 V switch to open
devx.ctrl_transfer(DevID, 0x40, 0x51, 37, 0, 0, 0, 100)
# set CHB GND switch to open
devx.ctrl_transfer(DevID, 0x40, 0x51, 38, 0, 0, 0, 100)
# open CHA voltage sense loop
devx.ctrl_transfer(DevID, 0x40, 0x51, 34, 0, 0, 0, 100)
# open CHB voltage sense loop
devx.ctrl_transfer(DevID, 0x40, 0x51, 39, 0, 0, 0, 100)