In cell attached mode, monitor resistance for break-in #181

Open
campagnola opened this issue Oct 5, 2023 · 1 comment
Open

In cell attached mode, monitor resistance for break-in #181

campagnola opened this issue Oct 5, 2023 · 1 comment
Assignees

Comments

@campagnola
Copy link
Contributor

No description provided.

@outofculture
Copy link
Collaborator

We cannot rely on just resistance, as a failed seal will also show a drop in resistance. This is therefore a need for a better capacitance calculation so that both can be used in concert to determine when a spontaneous break-in has occurred. Look at the neuroanalysis package for cell models that can provide testing of our analysis code.